Defining Elements of Roller Pump Occlusion in Cardiopulmonary
Bypass Surgery
SHOTA KATO1,2, SHOTA SOGABE1, JUN YOSHIOKA3, KAZUHIKO NAKADATE4,
HITOSHI KIJIMA1, YASUTOMO NOMURA2
1Department of Medical Technology and Clinical Engineering,
Gunma University of Health and Welfare,
191-1 Kawamagarimachi Maebashi Gunma 371-0823,
JAPAN
2Maebashi Institute of Technology Graduate School of Environment and Biotechnology,
460-1, Kamisadori Maebashi Gunma 371-0816,
JAPAN
3Sendai Red Cross Hospital Medical Technology Department,
Clinical Engineering Technology Division,
2-43-3, Hakugiyamahonchou Sendai Miyagi 982-8501,
JAPAN
4Department of Basic Science Educational and Research Center for Pharmacy,
Meiji Pharmaceutical University,
2-522-1Nojiri Kiyose Tokyo 204-8588,
JAPAN
Abstract: - In surgery, centrifugal pumps are used when safety and biocompatibility are priorities. On the other
hand, when considering operability and economy, roller pumps are used. The roller pump occlusion has to pay
attention to the long-time operation causing axis displacement; Our study carried out that perfusion temperature
executed the technology as a rule factor of the occlusion. To evaluate the manifestation of the circumference of
the occlusion, we used three kinds of different roller diameters and measured perfusion temperature (Pt) and the
electrical resistivity (Er), a pressure degree of the occlusion. Based on Japan Industrial Standard -T1603, we
observed the differences between the three pumps with degree of the occlusion which we were setting in the
same condition as a change of the occlusion by the progress at a time. Pt and Er repeated the up-down motion
in three pumps every 30-60 minutes. In addition, the occlusion extended the interval in progress at a time. The
pressure level of the sensor rose every 30 minutes and became unmeasurable afterward. This phenomenon
affects that perfusion temperature changes influence the blood viscosity and, we suppose that it influenced a
rise in para-blood temperature and it appeared in pressure change and Er of the occlusion. Therefore, control of
the Pt leads to the appropriate control of the roller pump and we will be able to carry out physiological
extracorporeal considering an indispensable element as SDGs.
Key-Words: - Occlusion, Perfusion Temperature (Pt), Electrical Resistivity (Er), Circumferential runout,
Pressure sensor, Blood cell destruction, Mechanical fracture.
Received: August 3, 2023. Revised: December 22, 2023. Accepted: January 28, 2024. Published: Arpil 1, 2024.
1 Introduction
As a characteristic of centrifugal pumps, the
perfusion amount may change due to low head and
changes in afterload, [1]. There are also reports that
there is no significant difference in blood damage
when comparing centrifugal pumps and roller
pumps used in heart-lung surgery, [2], [3].
WSEAS TRANSACTIONS on BIOLOGY and BIOMEDICINE
DOI: 10.37394/23208.2024.21.15
Shota Kato, Shota Sogabe, Jun Yoshioka,
Kazuhiko Nakadate, Hitoshi Kijima, Yasutomo Nomura
E-ISSN: 2224-2902
148
Volume 21, 2024
Roller pumps are easy to operate, but the drawback
is that the occlusion cannot be adjusted during
surgery. Furthermore, there is a concern that red
blood cells may be destroyed due to changes in the
occlusion. In the world of blood rheology, it is said
that a decrease in the negative charge (zeta
potential) contained in sialic acid in the red blood
cell membrane and changes in the morphology of
the outer membrane wall increase the aggregation of
blood cells, [4], [5], [6]. Therefore, as a previous
study, a method has been reported in which the
occlusion of a roller pump is converted into an
electrical resistance value (Er: Electrical Resistance)
and then quantified intermittently, [7].
In this study, we focused on the behaviors of Er
due to temperature changes (Pt : Perfusion
Temperature) to continuously monitor changes in
the occlusion during roller pump operation using
simulated blood.
2 Materials and Methods
2.1 Experimental Equipment
Three types of pumps with different roller diameters
are available : Terumo CV-8000 (roller diameter 42
mm), Technowood BP-150 (roller diameter 22 mm),
and Livanova S5 (roller diameter 31 mm). used in the
experiment (Figure 1).
Fig. 1: Experimental used three roller parts
diameters
For the blood circuit, a roller pump was set with a
Mera Excel line H 3/8 inch tube (MERA Corp.,
Tokyo, Japan), inner diameter 9.5 mm, outer
diameter 14.3 mm, Shore hardness A70 degrees, and
tensile strength 13.4 MPa. The experimental
environment temperature was set to 25 (±2) °C, and
simulated blood mixed with glycerin and
physiological saline (0.9% wt) and 40% glycerin
aqueous solution (4.18 cP) were mixed with human
blood at 25 °C (37 °C). The same viscosity) was
prepared, [8], [9].
2. 2 How to Measure Controlling Factors
In addition, to confirm the change in perfusion
amount when the solution was circulated through the
tube, the pressure closure degree of CV-8000,
BP-150, and S5 was set to 6 drops/min and 13
drops/min, and the maximum average value of five
flow rate displays (ultrasonic flow meter and
tachometer) was evaluated. When we compared the
perfusion rates between the ultrasonic flowmeter and
the tachometer, we found that the difference between
the occlusion rate of 13 drops/min was smaller than
the occlusion rate of 6 drops/min, so we decided to
use the occlusion rate of 13 drops/min as the standard.
The experiment was conducted (Table 1).
The experimental outline was as follows
JIS-T1603: the occlusion of each pump was set to 13
drops/min, the perfusion fluid volume was 500 mL,
the rotation speed was 97 rpm, the perfusion time
was 180 minutes, and the pumps were moved every
30 minutes to an arbitrary point A (Figure 2), and the
electrical resistance value Er and perfusion
temperature Pt were sequentially measured under the
same conditions, [10].
Table. 1 Volumetric flow rate of rotary flowmeter (R)
and ultrasonic flowmeter (U) with occlusion
of 6 drops and 13 drops
6 drops/min 97 rpm 25
Time[min]
0
30
60
CV-8000
R
2480
2488
2492
U
2718
2796
2804
BP
-150
R
2580
2580
2580
U
2634
2654
2652
S5
R
2520
2520
2520
U
2548
2582
2582
6 drops/min 97 rpm 25
Time[min]
0
30
60
CV-8000
R
2480
2488
2496
U
2694
2704
2742
BP
-150
R
2580
2580
2580
U
2578
2576
2592
S5
R
2520
2520
2520
U
2506
2534
2558
*R
Volumetric flow rate of rotary flowmeter
*U
Ultrasonic flowmeter
WSEAS TRANSACTIONS on BIOLOGY and BIOMEDICINE
DOI: 10.37394/23208.2024.21.15
Shota Kato, Shota Sogabe, Jun Yoshioka,
Kazuhiko Nakadate, Hitoshi Kijima, Yasutomo Nomura
E-ISSN: 2224-2902
149
Volume 21, 2024
Fig. 2: Electrical resistivity measurement circuit
2.3 Pressure Measurement using a Pressure
Sensor
A small pressure sensor PS-70KC M2: 7 MPa
(Kyowa Dengyo Co. Ltd, Tokyo, Japan) was fixed
to the area where the roller and raceway were
crimped (Figure 3), and pressure changes were
observed. WGI-400A-00 (Kyowa Dengyo Co. Ltd,
Tokyo, Japan) was used as the measuring instrument,
and measurements were made under the conditions
of a sampling frequency of 50 Hz, a bridge power
supply of DC2 V, 30 mA, and an output voltage of
10 V. In addition, the pressure value of the degree of
closure was measured every 30 minutes.
Fig. 3: Schema of pressure measurement sensor
install site for occlusion pressure
3 Results
3. 1 Changes in the occlusion due to Er and
Pt
(1) CV-8000
Figure 4 shows the average values of Er and Pt from
the start of CV-8000 perfusion to 180 min.
Furthermore, Er converged to 40 to 46 for a
roller with a diameter of 42 mm.
Er decreased by 0.5 30 min after the start,
and after that, it repeated up and down movements.
On the other hand, Pt showed a waveform with an
opposite tendency. The correlation coefficient of
regression analysis of Er and Pt was R2 =0.74
(p<0.05), which showed a strong correlation.
Fig. 4: Serial changes in Er and pt with CV-8000
(2) BP-150
Figure 5 shows the results for Er and Pt. The Er of
BP-150 converged to 37 to 44 kΩ.
From Figure 5, Er decreased by 0.6 30 min
after the start, and then repeatedly fluctuated up and
down. Pt showed a waveform with an opposite trend.
The correlation coefficient of regression analysis of
Er and Pt showed a significant correlation of R2
=0.61 (p<0.05).
Fig. 5: Serial changes in Er and pt with BP-150
(3) S5
The results for Er and Pt are shown in Figure 6. The
Er of S5 converged to 29 to 34 kΩ.
From Figure 6, Er decreased by 1 30 min
after the start, and then gradually decreased. The
waveform increased 150 minutes after the start but
decreased overall. On the contrary, Pt showed a
waveform with a tendency opposite to that of Er.
The correlation coefficient of regression analysis of
Er and Pt showed a significant correlation with
R2=0.66 (p<0.05).
WSEAS TRANSACTIONS on BIOLOGY and BIOMEDICINE
DOI: 10.37394/23208.2024.21.15
Shota Kato, Shota Sogabe, Jun Yoshioka,
Kazuhiko Nakadate, Hitoshi Kijima, Yasutomo Nomura
E-ISSN: 2224-2902
150
Volume 21, 2024
Fig. 6: Serial changes in Er and pt with S5
3.2 Changes in Pressure due to Changes in
the Occlusion
Before the start of the experiment, the B-150's roller
pump was set at a drop rate of 13 drops/min, and the
pressure value for the occlusion was 6 MPa, but it
increased to 7 MPa 30 minutes after the start, and 8
MPa after 60 minutes, pressure beyond the
measurement range was applied, making
measurement impossible.
4 Discussion
4.1 Degree of Occlusion Predicted from Pt
As a result of examining the correlation between Er
and Pt using pumps with different roller diameters,
the Er of the CV-8000 and BP-150 pumps converged
within the optimal occlusion, and S5 converged
around 30 to 34 kΩ. In S5, the value is lower than the
optimum occlusion, and it is necessary to mention the
internal structure of the pump related to pressure
closure.
However, in all pumps, a negative correlation
(inverse proportion) between Er and Pt was observed,
and a phenomenon in which the occlusion (electrical
resistance value) decreased as the perfusion
temperature Pt increased was observed. Especially in
BP-150, when the deviation of Pt is large, the
deviation of Er also becomes large. This is thought to
be due to the frictional heat caused by the roller
rubbing against the tube as Pt increases, which
activates the molecular motion of the perfused
substance, making the density sparse and reducing
the electrical resistance value Er. It is thought that the
increase in Pt lowers the viscosity of the substance,
indicating a state in which the electrical resistance
decreases, and decreases Er.
In the case of continuous perfusion, frictional
heat is accumulated when the roller contacts the tube,
but since the perfusion temperature Pt is below 50°C,
changes in the tube due to the viscosity of the
perfusate were ignored, [11].
4.2 Relationship between Changes in
Occlusion and Pressure Sensor
The pressure measuring device used in this study
only displayed discrete instantaneous values and did
not record waveforms, making it impossible to
perform a detailed analysis of pressure sensor
damage.
However, it is believed that the cause of the
sensor damage was that the drive shaft supporting the
rollers caused circumferential vibration, causing the
rollers to approach the raceway and apply high
pressure to the pressure sensor PS-70KC M2.
In addition, the pressure applied to the strain
gauge is considered to be vertical pressure and
pressure (shear stress) applied to the curved raceway
due to the rotation of the rollers acting on the
pressure sensor (Figure 7).
Fig. 7: Pressure to depended on a sensor and
Transformation of the tube
The pressure can be expressed by the following
equation, where N is the force with which the roller
pushes out the tube, is the pressure N0 that pushes
out the tube in an unpressurized state, and is the
fluid force acting on the roller, [12], (Figure 7)
N = N0 +  
= 
{(2r - 2wR) sinθ + R}
p
: Tube internal pressure
w
: Tube thickness
R
: Radius of roller r: Outside radius of tube cross
section
: Maximum contact angle between roller and
tube
: Width of the tube at any point in contact with
the roller
Based on the pressure measurement and
calculation formula of the roller pump, it is predicted
that the high pressure caused by circumferential
vibration will cause a major disturbance to the actual
WSEAS TRANSACTIONS on BIOLOGY and BIOMEDICINE
DOI: 10.37394/23208.2024.21.15
Shota Kato, Shota Sogabe, Jun Yoshioka,
Kazuhiko Nakadate, Hitoshi Kijima, Yasutomo Nomura
E-ISSN: 2224-2902
151
Volume 21, 2024
blood because it will increase the perfusion
temperature Pt.
4.3 Blood Cell Destruction Expected due to
Circumferential Vibration
When using a roller pump in a clinical setting, it is
necessary to pay attention to the fact that the
frictional heat generated by rubbing the tube changes
the plasma viscosity of the blood and changes the
internal structure of red blood cells, [13]. Red blood
cells are maintained at a constant distance by the
electrical repulsion of the zeta potential on the
membrane surface, [14]. However, the reduction in
the amount of sialic acid and the loss of protein is
thought to increase the rate of red blood cell
aggregate formation and reduce the repulsive force
between blood cells, thereby increasing blood
aggregation, [15], [16]. Therefore, the frictional heat
and pressure changes generated by the rotating
rollers can be expected to disrupt the red blood cell
membrane.
Therefore, the narrowing of the interval between
the occlusion due to circumferential vibration leads
to the collapse of blood cells, resulting in changes in
plasma viscosity due to heat, mechanical collapse,
and changes in blood viscosity due to mechanical
destruction. Blood cell destruction significantly
reduces biological homeostasis, so the occlusion of
the roller pump must be adjusted from time to time.
4.4 As a Quantitative Monitor of the
Occlusion
The need for quantitative monitoring of changes in
the occlusion is an important issue in blood injury.
JIS-B0621 sets standards for circumferential runout,
but the occlusion changes from the time of setting.
Therefore, continuous monitoring is necessary to
ensure proper blood delivery, and it is desirable to
adjust from time to time, [17].
As an indirect method for measuring the
occlusion, it is necessary to monitor the electrical
resistance value Er and simultaneously monitor the
perfusion temperature Pt. Since Er changes
depending on the viscosity of the perfused substance,
it is important to quantify the occlusion because
temperature, electrical resistance, and viscosity are
factors that interfere with each other due to the
temperature rise caused by circumferential vibration.
becomes.
However, when using Er in a clinical setting, it is
not practical because the current is higher than the
minimum sensing current of the human body.
Therefore, by installing thermometers before and
after the roller and continuously monitoring changes
in Pt, it becomes possible to continuously monitor the
degree of pressure closure. As a control factor for the
degree of pressure occlusion, perfusion temperature
Pt is a useful monitoring item and is essential.
5 Conclusion
In surgery, roller pumps are used when considering
operability and economy. The roller pump occlusion
has to pay attention to the long-time operation
causing axis displacement.
(1) Our study carried out that perfusion temperature
executed the technology as a rule factor of the
occlusion.
(2) To evaluate the manifestation of the
circumference of the occlusion, we used three kinds
of different roller diameters and measured perfusion
temperature (Pt) and the electrical resistivity (Er), a
pressure degree of the occlusion.
(3) Control the degree of pressure closure of roller
pumps used in heart-lung machines, it was
understood that the
degree of pressure closure changes due to many
factors, such as circumferential runout and
temperature changes.
(4) This study concludes that the degree of pressure
occlusion changes and that relative monitoring of
perfusion temperature is necessary.
(5) More detailed analysis and experiments are
required, but since each console is expensive, we
plan to confirm this in the future.
(6) Furthermore, we plan to compare not only the
roller pump but also the centrifugal pump in terms of
volumetric flow rate and blood cell morphology.
References:
[1] Glenn P. Gravlee, Richard F. Davis, Alfre H.
Stammers, Ross M. Ungerleider,
Cardiopulmonary Bypass: Principles and
Practice, 3rd Edition, Lippincott Willams and
Wilkins, 2007.
[2] J. W. Mulholland?J. C. Shelton?X. Y. Luo:
Blood flow and damage by the roller pumps
during cardiopulmonary bypass. Journal of
Fluid and Structures, Vol. 20, Issue 1,
pp129-140, 2005,
https://doi.org/10.1016/j.jfluidstructs.2004.10.
008.
[3] H Demirtas, E Iriz, CY Demirtas, D Erer, L
Oktar, A Yener, M Arslan, Investigating the
Effects of Two Different Pump Heads
(Centrifugal vs. Roller Pump) on
Hematological and Immunological
WSEAS TRANSACTIONS on BIOLOGY and BIOMEDICINE
DOI: 10.37394/23208.2024.21.15
Shota Kato, Shota Sogabe, Jun Yoshioka,
Kazuhiko Nakadate, Hitoshi Kijima, Yasutomo Nomura
E-ISSN: 2224-2902
152
Volume 21, 2024
Mechanisms, Nigerian Journal of Clinical
Practice, Vol. 21(7), pp.847-853, 2018.
https://doi.org/10.4103/njcp.njcp_427_16.
[4] PMD Act: Pharmaceutical and Medical Device
Act, Report on the results of the 3rd Meeting
on Measures for the Safe Use of
Pharmaceuticals, Medical Devices, and
Regenerative Medical Products, 2015,
[Online].
https://www.pmda.go.jp/files/000210376.pdf,
(Access Date: March 22, 2024).
[5] Nitesh Kumar, Numerical Investigation on the
effect of blood pressure on Wall Shear Stress
and Vorticity, WSEAS Transaction on Biology
and Biomedicine, Vol. 18, pp63-65, 2021,
https://doi.org/10.37394/23208.2021.18.7.
[6] Shota Kato, Tadashi Handa, Jun Yoshioka,
Kazuhiko Nakadate, Yasutomo Nomura and
Hitoshi Kijima, Evaluation of Blood Cells
Destruction by Measuring Occlusion Distance,
WSEAS Transactions on Biology on
Biomedicine, Vol. 20, pp. 313-320, 2023.
https://doi.org/10.37394/23208.2023.20.32.
[7] Naoki Momose, Asami Kitamura and Noriaki
Ono among other things, Electric
Measurement of Occlusion degree in a Roller
Pump-driven Circuit, Japanese Journal of
Extra-Corporeal Technology, Vol.30, No.1, pp.
22-25 2003.
[8] Wataru Hijikata, Jun Rao, Shodai Abe, Setuo
Takatani, Tadahiko Shinshi, Sensorless
Viscosity Measurement in a
Magnetically-Levitated Rotary Blood Pump,
Artificial Organs, Vol. 39(7), pp. 559-568,
2016, https://doi.org/10.1111/aor.12440.
[9] Kun-Lin Lee, Glen Kowach, Fang Li and Ioana
Voiculescu, Liquid Viscosity Sensor Using a
Surface Acoustic Wave Device for Medical
Applications Including Blood and Plasma,
Sensor, Vol. 23(13), 5911, 2023,
https://doi.org/10.3390/s23135911.
[10] JIS-T1603: Cardiopulmonary machine electric
blood pump, Electric motor driven blood pump
for cardiopulmonary bypass, 1995.
[11] Parisa Khoshnoud and Nidai Abu-Zahra,
Properties of rigid polyvinyl chloride foam
composites reinforced with different shape
fillers, Journal of Thermoplastic Composite
Materials, Vol. 30(11), pp. 1541-1559, 2017,
https://doi.org/10.1177/0892705716646417.
[12] S. Mukhopadhyay, G.C. Layek, Sk. A. Samad,
Study of MHD boundary layer flow over a
heated stretching sheet with variable viscosity,
International Journal of Heat and Mass
Transfer, Vo. 48(21-22), pp. 4460-4466, 2005,
https://doi.org/10.1016/j.ijheatmasstransfer.20
05.05.027..
[13] Hideyuki Tsukagoshi, Seiichi Nozaki, Kouichi
Nishizawa and Ato Kitagawa, Development
and Motion Analysis of the Inscribed
Pinch-roller Type Fluid Power Motor Driven
by Tap Water Pressure, Transactions of the
Japan Fluid Power System Society, Vol. 35(6),
pp. 109-116, 2004,
https://doi.org/10.5739/jfps.35.109.
[14] Artur S. Bartosik, Effect of High Solid
Concentration on Friction in a Transitionaland
Turbulent Flow of Bioliquid Suspension,
WSEAS Transaction on Fluid Mechanics, Vol.
18, pp.10-23, 2023,
https://doi.org/10.37394/232013.2023.18.2.
[15] Kung-Ming Jan, Shu Chien, Role of Surface
Electric Charge in Red Blood Cell Interactions,
The Journal of general physiology, Vol. 61(5),
pp. 638-654, 1973,
https://doi.org/10.1085/jgp.61.5.638.
[16] Takahiko Aoki, A Comprehensive Review of
Our Current Understanding of Red Blood Cell
(RBC) Glycoproteins, Membranes, Vol. 7(4),
pp. 56-74, 2017,
https://doi.org/10.3390/membranes7040056.
[17] ISO 1101: 1983(MOD), Maching Symbols,
[Online].
https://www.scribd.com/doc/131937901/ISO-
1101-1983-E-Maching-Symbols (Accessed
Date: March 12, 2024).
WSEAS TRANSACTIONS on BIOLOGY and BIOMEDICINE
DOI: 10.37394/23208.2024.21.15
Shota Kato, Shota Sogabe, Jun Yoshioka,
Kazuhiko Nakadate, Hitoshi Kijima, Yasutomo Nomura
E-ISSN: 2224-2902
153
Volume 21, 2024
Contribution of Individual Authors to the
Creation of a Scientific Article (Ghostwriting
Policy)
- Shota Kato conducted all of the experiments and
wrote the submitted paper.
- Shota Sogabe conducted an experiment on the
volumetric flow rate of ultrasound.
- Jun Yoshioka created a blood cell preservation
solution.
- Kazuhiko Nakadate provided mathematical advice.
- Hitoshi Kijima gave some electric advice.
- Yasutomo Nomura provided mathematical analysis
and advice throughout the experiments.
Sources of Funding for Research Presented in a
Scientific Article or Scientific Article Itself
No funding was received for this study.
Conflict of Interest
The authors declare no conflicts of interest relevant
to the content of this article.
Creative Commons Attribution License 4.0
(Attribution 4.0 International, CC BY 4.0)
This article is published under the terms of the
Creative Commons Attribution License 4.0
https://creativecommons.org/licenses/by/4.0/deed.en
_US
WSEAS TRANSACTIONS on BIOLOGY and BIOMEDICINE
DOI: 10.37394/23208.2024.21.15
Shota Kato, Shota Sogabe, Jun Yoshioka,
Kazuhiko Nakadate, Hitoshi Kijima, Yasutomo Nomura
E-ISSN: 2224-2902
154
Volume 21, 2024